SP295C SoundPro Audio Analyzer Contractor Version
Overview
The SP295 SoundPro Audio Analyzer Contractor Version is now available with enhanced capabilities to provide the audio analysis tools you need to install, calibrate, or analyze commercial sound venues such as churches, stadiums, meeting rooms, live stages, clubs, etc.
The Auto Time Delay test has been added to the SP295C to test the time delay from each speaker in a home theater surround sound system, as compared to the left channel speaker. It is designed to be used with the DAG5161 SurroundPro Digital Audio Generator. The DAG5161 is used to generate the "Auto Time Delay" signal. This signal sends out pulses to each of the five or six main channels, spaced 100ms apart. The SP295C's microphone detects these pulses and calculates the times between them to determine the time synchronization for each of the speakers.
Features
Specifications
| SPL | |
| Sound Level Meter: True RMS measurement of ambient sound energy levels | |
| Inputs: | Microphone and line |
| Display: | Digital and analog with peak hold |
| Accuracy: | ±0.3 dB SPL, steady state |
| Resolution & Range: | 0.1 dB; 35 - 125 dBA with supplied microphone |
| Weighting networks: | ANSI Type 1 |
| A, B, and C-weighting, Flat, 1 and 1/3 octave filters | |
| Averaging: | ANSI Type 1 |
| Slow (1000 mS exponential decay, time-averaged) | |
| Fast (125 mS exponential decay, time-averaged) | |
| Impulse (35 mS exponential decay) | |
| Peak | |
| LEQ (10 Sec. 1, 5, 10, or 15 min. 1, 8, or 24 hour or manual) | |
| Sound Study Graph: Graphically displays SPL level over a period of time | |
| Inputs: | Microphone and line |
| Averaging: | Peak or average |
| Weighting: | ANSI Type 1: A, B, and C-weighting, Flat, 1 and 1/3 octave filters |
| Time Range: | 1, 5, 10, and 15 min., 1, 2, 8, and 24 hours |
| Resolution: | Time range divided into 120 equal time segments |
| Range: | 35-125 dBA with supplied microphone in 2 selectable ranges |
| ACOUSTICS | |
| Real-Time Analyzer: Uses ANSI Class 1 octave-band or 1/3 octave-band filters to show relative sound levels by frequency band | |
| Inputs: | Microphone and line |
| Band Filters: | Selectable 1 or 1/3 octave |
| Filter Accuracy: | ± 0.20 dBu at band center frequency |
| Frequency Range: | Full (20 Hz to 20 kHz) or Low (10 Hx to 332 Hz) |
| Averaging: | Exponential Decay (1, 3, 6, 10, 30, 60 sec.), Equal-weighted Time (Avg), and Peak Hold |
| Display: | Graphic, digital SPL, digital SPL of band selected by cursor |
| Range & Resolution: | 17 or 35 dB window (0.5 dB/pixel or 1.0/pixel) within a 25 to 170 dB range, adjustable in 5 dB steps and 2 input ranges |
| Level Accuracy: | ± 1.5 dB with supplied microphone |
| Weighting: | Flat, A-weighted, C-weighted, Cinema X (ANSI PH22.202M-1984) |
| FFT Analyzer: Plots sound energy levels vs. octave band filtered spectrum | |
| Inputs: | Microphone and line |
| Band Filters: | Full octave, 1/3 octave, 1/6 octave, 1/12 octave |
| Filter Accuracy: | ± 0.20 dBu at band center frequency |
| Averaging: | Exponential Decay (1, 3, 6, 10, 30, 60 sec.), Equal-weighted Time (Avg), and Peak Hold |
| Display: | Graphic, digital SPL, digital SPL of band selected by cursor |
| Range & Resolution: | 17 or 35 dB window (0.5 dB/pixel or 1.0/pixel) within a 25 to 170 dB range, adjustable in 5 dB steps and 2 input ranges |
| Level Accuracy: | ± 1.5 dB with supplied microphone |
| Freq. discrimination: | ± 1% |
| Weighting: | Flat, A-weighted, C-weighted, Cinema X (ANSI PH22.202M-1984) |
| Energy Time Graph: Real time plot of impulse sound decay energy vs. time or sound propagation distance | |
| Inputs: | Microphone and line |
| Graph y-axis unit: | ms (milliseconds), ft (feet), m (meters) |
| Maximum time: | 7680 msec. |
| Maximum distance: | 8678 ft, 2641 m |
| Display: | Graphical display in real-time or freeze, measuring cursor displays time/distance and dB level, delay time to initial pulse, RT0 marker RT10 — RT60 reverberation times using cursors |
| Display Accuracy: | ± 1 pixel |
| Weighting: | ANSI A, B, C, octave, and 1/3 octave-band |
| Reverb Decay Time: Measures reverberation decay time and RT60 using pink noise. When a 60 dB dynamic range is not available, RT60 is linearly extrapolated from available dynamic range. | |
| Weighting: | A-Weighted, B-Weighted, C-Weighted, Flat, 1 and 1/3 octave band |
| Range: | 5 msec to 14.5 seconds, RT60 time to 11 sec. |
| Ambient Noise: | Ambient noise level must be at least 26 dB ower than test signal |
| Accuracy: | ± 2% full signal |
| Resolution: | 10 msec. |
| RT60 Time: | Extrapolated from Decay Time measurement |
| Noise Criteria: Implements ANSI S12.2-1995 for comparison of background noise level in nine octave bands to standard noise criterion curves | |
| Inputs: | Microphone and line |
| Displays: | NC curve, limiting band, SPL of selected octave band, Speech Interference Level |
| TECHBENCH | Level Meter/Frequency Counter |
| Level Meters: Display types are digital left and right line input levels. Analog bar graph and digital left and right line input VU levels. Peak Program Material line input levels. | |
| Inputs: | Line inputs. |
| Reference Level: | -30 to +20 dBu (VU/PPM meter) |
| Range: | -95 dBu to +40 dBu, auto or manual range |
| Units: | dBu (0 dB referenced to 0.775 Vrms), v (0 dB referenced to 1.00 Vrms), Vave (average voltage), Vrms (true RMS voltage calculation), Vp-p (true peak-to-peak algorithm) |
| Accuracy: | ± 0.5 dB, 20 Hz to 20 kHz, +40 dBu to -85 dBu |
| Frequency Counter: | |
| Inputs: | Line inputs. |
| Display: | L & R displayed independently |
| Range: | 16 Hz to 50 kHz |
| Accuracy: | ± 1 Hz mixed font |
| Signal Generator: Sine wave, square wave, white noise, pink noise (full, 1 or 1/3 octave band limited), Polarity waveform (square wave top and sine wave bottom or signal tracing) | |
| Distortion: | <0.1% sine wave output |
| Output Frequency Range: | Sine wave 1 Hz to 22 kHz, square waves 1 Hz to 1.6 kHz in 1 Hz, 1/3 octave or 1 octave steps |
| Output Level: | Sine wave, -78 dBu to +17 dBu |
| Square wave, -78 dBu to +17 dBu | |
| White noise, -78 dBu to +2 dBu | |
| Pink noise, -78 dBu to 0 dBu | |
| Level Accuracy: | ± 0.05 dBu @1000 Hz, ± 0.10 dBu 16 Hz to 20 kHz |
| Filter Accuracy: | Pink noise filters <0.05 dB, 20 Hz to 19 kHz, <1 dB at 20 kHz |
| Resolution: | Fine: 0.05 dB from -9.95 to +9.95, 0.1 dB all others |
| Coarse: 1.5 dB | |
| Display: | Signal type, frequency, level set, output jack, level measurement units, actual signal level at output (compensated for loading) |
| Impedance Meter: | Measures impedance of external device at any frequency from 20 Hz to 20 kHz, in 1/3 octave steps. |
| Impedance Range: | 1 ohm to 8000 ohms, 20 Hz to 20 kHz |
| Impedance Display: | Test frequency, impedance, equivalent Wattage for 25V, 50V, 70V, or 100V at selected frequency |
| Impedance | 2 ohms to 50 ohms: ± 2%, 50 Hz - 16 kHz; |
| Accuracy: | 1 ohms to 8000 ohms: ± 10%, 20 Hz - 20 kHz; |
| Sweeps | |
| Sine wave sweep signal: Smooth sweep between two selected frequencies over selected 10 to 99 seconds time duration. | |
| Amplitude sweep signal: Plots average level in each 1/3 or 1/12 octave band within selected sine wave sweep band. | |
| Impedance sweep signal: Plots average impedance in each 1/3 or 1/12 octave band within selected sine wave sweep band. | |
| Frequency Range: | 20 Hz to 20 kHz |
| Impedance Range: | 20 ohms to 8000 ohms |
| Display: | Impedance & Amplitude - graphic, Sine wave - digital |
| Resolution: | 1/3 octave - 30 points, 1/12 octave - 120 points |
| Accuracy: | ± 0.5 dB or 1 pixel mixed fonts |
| Distortion Meter: Compares level of harmonics and noise in input signal to the applied fundamental test frequency. | |
| Inputs: | Microphone and line |
| Display: | Digital and log scale bar graph |
| Display Range: | 0.02% to 50 % THD+N |
| Accuracy: | ± 5% |
| Test Frequencies: | 63 Hz, 125 Hz, 250 Hz, 500 Hz, 1 kHz, 2 kHz, 4 kHz |
| Signal-to-Noise Ratio: Samples the device's full-scale signal output and its noise floor and computes the device's S/N ratio. | |
| Inputs: | Microphone and line |
| Display: | Simultaneous display of left and right channel |
| Range: | 120 dB; minimum noise level -80 dBu |
| maximum signal level +40 dBu | |
| TESTS | |
| Auto HT Tests: Controls DAG5161 Digital Audio Generator for automated calibration procedures | |
| Tests: | Speaker Level Balance, Speaker Polarity, Speaker Time Delay, Reverb Decay, Early Reflections, Ambient Noise, LFE Frequency Response, System Frequency Response |
| Auto Time Delay: Uses test signal from DAG5161 for time delay calibration | |
| Tests: | Time Delay Speaker Calibration |
| Polarity Tests: Uses DSP analysis to determine the polarity of the applied test pulse vs. the signal picked up by mic or applied to the line input | |
| Tests: | Speaker, Mic, or Equipment (audio signal chain polarity) |
| Range: | Speaker test: 0 to 500 ft, speaker to SoundPro |
| Cable Test: Applies digital wiremap test to check for opens, shorts, crosses, and analog transmission test to compare transmission loss at 1 kHz and 20 kHz | |
| Connector Types: | RCA, XLR, 1/4" balanced and unbalanced, XLR to 1/4" |
| Display: | Graphical wiremap, attenuation loss |
| UTILITIES | |
| Audio Scope: | |
| Sample Rate: | 48 kHz |
| Displays: | mV vs. time (line in), mPa vs. time (mic in), or X-Y phase dislay |
| Display Amplitude: | 15.6 - 1000 mV/div (line in), 7.8 - 500 mPa/div (mic in) |
| Accuracy: | ± one pixel of display, 20 Hz to 22 kHz |
| Monitor Amplifier: | |
| Frequency response: | Signal sampled at 48 kHz 16-bit |
| Amplification: | Line input: variable 0-22 dB gain with selectable 40dB amp or 40 dB pad, 60 dB gain maximum |
| Mic input: variable 0-22 dB gain | |
| Load impedance: | 32 ohms minimum |
| Inputs: | Microphone and stereo line |
| Outputs: | Summed into mono, line outputs (XLR, 1/4", RCA phono jack), built-in speaker, headphone jack. |
| Output Level: | Variable from -79 to 0 dB |
| Phantom Power Test: Measures DC voltage on XLR "+" and "-" pins (pins 2 & 3) | |
| Accuracy: | ± 00.2 volts |
| GENERAL | |
| Display: | 64 X 128 pixel super-twist LCD, w/LED backlight |
| Size: | 10 3/8" x 6" x 3 3/4" (HWD) |
| Weight: | 3 lbs. 5 oz. (1502 g.) |
| Battery: | 6 volt rechargeable lead-acid cell provides 2-4 hours of service |
| Battery Life Display: | Accurate to ± 0.02 Vdc |
| AC Adapter/Charger: | 6 V @ 800 mZ |
| Non-Volatile Memory: | 40 graph data storage locations |